| Alco-Sensor, Alco-Sensor III |
| Symptom |
Cause |
Test Procedure |
Solution |
| Display .888” |
Low battery
|
Press and hold READ button, observe display
|
Replace with fresh Alkaline battery |
| Display has missing digit(s) and/or missing segment(s) in digit(s) |
IC(s) on circuit board have become unseated |
Press and hold READ button, observe display |
Send to AlcoPro for evaluation |
| Calibration screw does not work |
Using Previous Version calibration procedure for New Version instrument |
Mouthpiece ejector button = New Version Instrument |
Use correct calibration instruction. Click here |
| Calibration screw does not work |
Calibration Screw is broken |
In correct calibration mode, numbers do not change when turning screw. |
|
| Alco-Sensor IV (Any dot color) Back to Top |
| Symptom |
Cause |
Test Procedure |
Solution |
| Display “BAT” |
Low battery |
Insert mouthpiece, observe display |
Replace with fresh Alkaline battery |
| Keeps recycling back to temperature or temperature stays on and will not go to “TEST” |
Low battery |
Insert mouthpiece, observe display |
Replace with fresh Alkaline battery |
| Display “V004,” do not hear ‘click’ when taking sample |
Unit not taking sample |
Insert mouthpiece, press MANUAL button when “TEST” flashes, observe display |
Send to AlcoPro for evaluation |
| Unit will not perform test – will only display “V014” |
Voided sample analysis |
Insert mouthpiece, blow into mouthpiece when “TEST” flashes, observe display |
Recalibrate unit. If recalibration does not work,
Send to AlcoPro for evaluation |
| Display flashing “MAN” instead of “TEST”, unit will operate only when using ‘Manual’ button – will not auto sample |
Breath Thermistor is not operating |
Insert mouthpiece, observe display |
Send to AlcoPro for evaluation |
Displays “+” and “NoGo” without a sample being blown
May, intermittently, take sample normally |
Breath Thermistor is not operating correctly |
Insert mouthpiece, observe display |
Send to AlcoPro for evaluation |
| Not holding calibration or peak time is over 1 minute |
Fuel cell may be weak |
Perform accuracy check, noting result and time it takes to display result |
Send to AlcoPro for evaluation |
| Alco-Sensor IV w/Memory black dot Back to Top |
| Symptom |
Cause |
Test Procedure |
Solution |
Display “Mem Full” |
Internal memory full |
Insert mouthpiece, observe display |
|
| Printing incorrectly |
Printer setting |
Perform test and print result |
|
Date/Time incorrect on printouts |
Memory error |
Perform test, observe Time/Date on initial display AND printout
|
|
| RBT IV Back to Top |
| Symptom |
Cause |
Test Procedure |
Solution |
Incorrect Time/Date |
Time /Date not updated
|
Turn on RBT printer, observe Time/Date
|
|
| Will not print
|
Low charge on 12volt battery
|
Perform “Last Test” on printer
|
Plug AC charger into RBTIV case and electrical outlet overnight. Replace battery if necessary. |
| RBT printer displays “Memory Full”
|
RBT IV memory is full |
Turn RBT printer on, press START button to begin a test sequence, observe display |
Erase RBT memory.
See Instructions |
| RBT printer displays “Dump and Erase” |
RBT IV memory is almost full |
Turn RBT printer on, press START button to begin a test sequence, observe display
|
|
| Alco-Sensor FST Back to Top |
| Symptom |
Cause |
Test Procedure |
Solution |
Display “bAt” |
Low battery
|
Turn instrument on, observe display
|
Replace batteries with two
fresh AA Alkaline batteries |
| Does not sample Dry Gas automatically when doing Accuracy Check
|
Using 1.5 LPM regulator
|
Note 1.5 LPM label on regulator
|
Press ON button for Manual Sample when performing Accuracy Check.
Click here for instructions.
(Only regulators marked as 6 LPM flow rate will produce automatic sample for Accuracy Checks.) |
| Instrument does not turn on when pressing ON button
|
ON button broken
|
Press and hold ON button with trigger finger, observe display
|
|
| Tank and Regulator Back to Top |
Symptom |
Cause |
Test
Procedure |
Solution |
| Regulator gauge reads 900 psi or greater |
Full gas tank |
Observe regulator gauge |
Gas tank is full and ready for use
|
| |
|
|
|